Description
The AMETEK LEVEL MATE III Level System uses a stainless steel transmitter as standard. A series of perforations in the bottom of the unit protects the sensing port from either clogging or damage. The same snub-nosed end can also unscrew, allowing for easy calibration checks.
The customized LEVEL MATE III meter displays data from the transmitter, provides excitation voltage and transmits an analog signal for remote recording, alarm or control functions. It is easy to calibrate by front panel adjustments. Its programmable relay permits set and reset programming. The large, easy-to-read LED display gives local indication while the optional isolated analog output can interface with ancillary display, recording or control devices. Combined into one convenient package, the AMETEK LEVEL MATE III Level System provides you with the most versatile level measurement system on the market!
Features
Two-year factory warranty on the system
Intrinsically safe (Model 575S, 575P and 675)
Solid-state silicon sensor with milliamp output
Rugged stainless steel sensor housing – excellent environmental protection
Excellent media compatibility
Sensor vented to the atmosphere through the surface end of the cable
Microprocessor-controlled meter for quick, easy setup and calibration
Large dual line 6 digit display, 0.60” & 0.46”
Selectable setup via front keys
Programmable displays and function keys
Meter can be calibrated for inches of water, feet of water, meters of water, or any desired engineering units
Multi pump alternation control
Options Include
Lightning and secondary surge protectors
4 relay expansion module
Meter copy programming cable
4-20 mA analog output from meter
USB serial communication kit for configuring, monitoring and datalogging from a PC using DDMC MeterView software
Applications
Complete easy-to-use system – all in one package
Variety of reliable stainless steel sensors
Microprocessor-controlled digital display
Setpoints and signal outputs for remote monitoring, recording and control
Continuous measurement and display
Simple installation, virtually no maintenance
Low system pricing
Specs
Ranges:
Feet of Water: 0 to 14, 0 to 35, 0 to 69, 0 to 138, 0 to 230, 0 to 345, 0 to 460, 0 to 690
Meters of Water: 0 to 4, 0 to 10, 0 to 21, 0 to 42, 0 to 70, 0 to 105, 0 to 140, 0 to 210
Transmitter Temperature Ranges:
Max. Operating:
-40° to 82°C (-40° to 180°F); media must not freeze
Compensated:
-5° to 55° C (23° to 130°F)
Temp. Effects:
Typical ±1% output change for ±25° C temp. change within compensated range when calibrated at 25°C
Accuracy:
±0.25% full scale (including linearity, hysteresis, repeatability) B.F.S.L.
Transmitter Output:
4 to 20mADC, limited to 30 mADC
Transmitter Power Supply:
24 VDC from meter
Transmitter Electrical Connection:
Attached #20 AWG polypropylene insulated and shielded conductors in a black polyurethane jacket. Unspliced lengths to 3000 ft. (1622 m)
Meter Housing Construction:
Fiberglass reinforced polyester
Transmitter Cable Color Code:
RED ( ) excitation
BLACK (-) excitation,
GREEN (caseground).
Transmitter Construction:
Diaphragm: Type 316L stainless steel
Housing Type: 316 stainless steel
Nut/Washer Type: 316 stainless steel
Cable Grommet: Viton
Meter Display:
LED Display
Upper Display: 0.60” (15mm) high
Lower Display: 0.46” (12mm) high
Both displays 6 Digits (-99999 to 99999)
Meter Input Voltage:
85 to 265 VAC @ 50/60 Hz, 90-265 VDC, 20 W max
Meter Ranges:
Calibrated to transmitter range in psi, inches, feet, meters of water, or other desired engineering units -9999 to 99999.
Meter Temperature Ranges:
Operating: -40° to 65° C (-40° to 150°F)
Storage: -40° to 85°C (-40° to 185°F)
Meter Output:
Isolated 4-20 mA.
While the output is nominally 4-20 mA, the signal will accurately accommodate under and over ranges from 1 to 23 mA.
-
Level-Mate-III-Manual.pdf
3 KB
美国
-
DDMC-Datasheet.pdf
2 KB
美国
-
LevelMateIII-A_EDO_7-11-101_Issue_1.pdf
1270 KB
美国
-
WWSVP-A_2-13-110_Iss_4.pdf
1109 KB
美国